Detection efficiency of NaI(Tl) detector based on the fabricated calibration of HPGe detector
Abstract
Determination of the full-energy peak efficiency of a detection system is significant in the activity calculation of the measured radioactive samples. In this work, we present an experimental method to determine the absolute efficiency calibration of a NaI(Tl) detector, considering that a standard source of interest is unavailable and by using the known specific activity of a standard sample measured using a HPGe detector. The efficiency of a gamma-ray spectrometer that consists of a coaxial HPGe detector is calculated using Canberra ISOCS/LabSOCS software and a fabricated calibration source. To validate our method, environmental samples (rock and soil samples) were analyzed on both the detectors. The obtained activity concentrations were checked by means of the IAEA proficiency test procedure. The performance criteria evaluation results were found to be ‘Acceptable’ for all the analytical determinations of all …